The AIP9620TX and AIP9620X leakage current testers perform leakage current testing using a human body simulation network module and insulation resistance, and are widely used in various electrical safety testing applications.
■ Human body simulation test: Built-in human body simulation network module that meets the requirements for human body contact leakage current testing.
■ Dual-frequency output: Supports AC 50Hz / 60Hz testing, adaptable to different power grid standards worldwide.
■ Intelligent judgment: Upper and lower limits can be preset, automatic pass/fail judgment is provided, and audible and visual alarms are provided.
■ High-definition display: The LCD high-definition screen displays test data in real time, and parameter settings are intuitive and convenient.
■ Dual-mode result display: Test results can be switched between resistance and voltage display to meet different standard requirements.
■ Widely applicable: Suitable for household appliances, lighting, wires and cables, electronic components, power tools, and laboratories.
■ High-precision CPU control: Employs high-speed CPU intelligent control technology to achieve stable and accurate leakage current measurement.
■ Multiple safety protections: Equipped with automatic overvoltage and overcurrent protection mechanisms to effectively ensure equipment and operational safety.
■ Compliant with international safety standards: Meets electrical safety testing specifications such as IEC 60950, UL 60950, IEC 60335-1, UL 60335-1, and GB 4706.1.
■ Meets metrological verification requirements: Complies with safety metrological verification procedures such as JJG 795-2004 and JJG 690-2003, and meets calibration and quality system requirements.
| Model | AIP9620TX | AIP9620X |
| Input voltage | 220Vac±10% | |
| Input frequency | 50/60Hz±5% | |
| Fuse | Fuse 0.5A/250V | |
| Leakage current | ||
| Test input voltage | 30~450V AC three-phase | 30~ 300 V AC single-phase |
| Test input current | 20A AC MAX | 20A AC MAX |
| Voltage display range | 30~450V AC | 30~ 300 V AC |
| Voltage display resolution | 1V | 1V |
| Voltage display accuracy | ± (2% of the set value + 3V) | ± (2% of the set value + 3V) |
| Leakage current range | 50-6000uA | 50-6000uA |
| Leakage current resolution | 1uA | 1uA |
| Leakage current accuracy | ± (2% of the reading + 30uA) | ± (2% of the reading + 30uA) |
| Built-in human body network | GB4706.1-2005 version requirements | GB4706.1-2005 version requirements |
| Test time range | 1.0-2.0-999.9s | 1.0-2.0-999.9s |
| Test time resolution | 0.1s, set to test only live-wire leakage for 1 second. | 0.1s, set to test only live-wire leakage for 1 second. |
| Functional parameters | ||
| Display Interface | 16*2 Blue Screen Character LCD Display | |
| Communication interface | RS232 (optional) | |
| Correction method | Software calibration can be performed by inputting parameters via the front panel. | |
| Work environment | 0-40℃, 20-80%RH | |
| External dimensions | 280mm (W) × 89mm (H) × 370 mm (D) | |
| Random accessories | Power cord ×1, test clips ×2, fuse ×1, instruction manual ×1 | |
Quick Selection Reference | ||||||
| Model | Test Project | Accuracy | ||||
AC Hi-pot (ACW) | DC Hi-pot (DCW) | Insulation resistance (IR) | Grounding resistance (GB) | Leakage current (LC) | ||
| AIP9620X | — | — | — | — | 300V, 20A | 2% |
| AIP9620TX | — | — | — | — | 450V, 20A | 2% |